Background:
Patients treated for sick sinus syndrome may have interatrial conduction disorder leading to atrial fibrillation.
Objectives:
This study was aimed to assess the influence of the atrial pacing site on interatrial and atrioventricular conduction as well as the percentage of ventricular pacing in patients with sick sinus syndrome implanted with atrioventricular pacemaker.
Material And Methods:
The study population: 96 patients (58 females, 38 males) aged 74.1 ± 11.8 years were divided in two groups: Group 1 (n = 44) with right atrial appendage pacing and group 2 (n = 52) with Bachmann's area pacing. We assessed the differences in atrioventricular conduction in sinus rhythm and atrial 60 and 90 bpm pacing, P-wave duration and percentage of ventricular pacing.
Results:
No differences in baseline P-wave duration in sinus rhythm between the groups (102.4 ± 17 ms vs. 104.1 ± 26 ms, p = ns.) were noted. Atrial pacing 60 bpm resulted in longer P-wave in group 1 vs. group 2 (138.3 ± 21 vs. 106.1 ± 15 ms, p < 0.01). The differences between atrioventricular conduction time during sinus rhythm and atrial pacing at 60 and 90 bpm were significantly longer in patients with right atrial appendage vs. Bachmann's pacing (44.1 ± 17 vs. 9.2 ± 7 ms p < 0.01 and 69.2 ± 31 vs. 21.4 ± 12 ms p < 0.05, respectively). The percentage of ventricular pacing was higher in group 1 (21 vs. 4%, p < 0.01).
Conclusions:
Bachmann's bundle pacing decreases interatrial and atrioventricular conduction delay. Moreover, the frequency-dependent atrioventricular conduction lengthening is much less pronounced during Bachmann's bundle pacing. Right atrial appendage pacing in sick sinus syndrome patients promotes a higher percentage of ventricular pacing.
Related Concept Videos
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...
Dysrhythmias I: Introduction
Mechanism of Cardiac Arrhythmias