Sheath placement in femoral artery during cardiac catheterization in children can influence pressure waveform
Masahiro Shiraishi1, Tomoaki Murakami1, Tomohiro Nawa1
1Department of Cardiology, Chiba Children's Hospital, 579-1, Heta-cho, Midori-ku, Chiba 266-0007, Japan.
Insights
Femoral arterial sheath placement during cardiac catheterization significantly decreases the d/a ratio, indicating enhanced aortic pressure wave reflection. This finding suggests that the procedure itself alters central aortic pressure waveforms in pediatric patients.
Area of Science:
- Cardiovascular Physiology
- Medical Device Impact
- Pediatric Cardiology
Background:
- External compression of the femoral artery is known to enhance pressure wave reflection.
- Cardiac catheterization procedures may inherently influence aortic pressure waveforms.
Purpose of the Study:
- To investigate the impact of femoral artery sheath placement on aortic pressure waveforms.
- To determine if cardiac catheterization influences central aortic pressure dynamics.
Main Methods:
- The study involved 21 pediatric patients undergoing cardiac catheterization with femoral artery sheath placement.
- Aortic pressure waveform changes were assessed using the b/a and d/a ratios from fingertip photoplethysmography.
- The d/a ratio signifies aortic pressure wave reflection, with lower values indicating enhancement.
Main Results:
- Femoral sheath placement did not significantly alter the b/a ratio (p=0.896), suggesting no change in aortic stiffness.
- A significant decrease in the d/a ratio was observed post-sheath placement (p=0.0003).
- This reduction in d/a ratio indicates enhanced aortic pressure wave reflection.
Conclusions:
- Placement of a femoral artery sheath enhances aortic pressure wave reflection.
- This procedural effect can alter the central aortic pressure waveform during cardiac catheterization.
Introduction:
It is reported that pressure wave reflection is enhanced by external compression of the femoral artery. Therefore, it is possible that cardiac catheterization itself can influence the aortic pressure waveform.
Aim:
The purpose of this study is to clarify the influence of sheath placement in a femoral artery on the pressure waveform.
Methods:
This study enrolled 21 pediatric patients (5.1±4.0years) who underwent cardiac catheterization. A sheath was placed in the femoral arteries of all patients. The change in the pressure waveform induced by the placement of the sheath was investigated using the b/a and d/a ratio of second derivative of a fingertip photoplethysmogram. A high b/a ratio means a stiff aorta and a low d/a ratio represents an enhancement of the aortic pressure wave reflection.
Results:
By the placement of the sheath in their femoral arteries, the b/a ratio was not influenced (sheath (-): -0.556±0.081 vs. sheath (+): -0.558±0.072; p=0.896). However, the d/a ratio was significantly decreased (-0.150±0.074 vs. -0.185±0.084; p=0.0003).
Conclusions:
The placement of the femoral arterial sheath enhances the pressure wave reflection and would lead to a change in the central aortic pressure waveform.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
05:36Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: