Optical Fiber Geometry: Accurate Measurement of Cladding Diameter
Matt Young1, Paul D Hale1, Steven E Mechels1
1National Institute of Standards and Technology, Boulder, CO 80303-3328.
Summary
Three new instruments accurately measure optical fiber cladding diameter to under 50 nm. These tools, including a micrometer for creating reference materials, enhance precision in fiber optic metrology.
Area of Science:
- Metrology
- Optical Engineering
- Materials Science
Background:
- Accurate measurement of optical fiber cladding diameter is critical for telecommunications and data transmission.
- Existing metrology techniques may lack the required precision for advanced fiber optic applications.
Purpose of the Study:
- To develop and validate novel instruments for high-accuracy optical fiber cladding diameter measurement.
- To establish traceable measurement standards for optical fiber dimensions.
Main Methods:
- Development of a contact micrometer for direct diameter measurement.
- Implementation of a scanning confocal microscope for non-contact surface profiling.
- Utilisation of a white-light interference microscope for high-resolution topographical analysis.
Main Results:
- All three developed instruments achieved an estimated uncertainty (3 standard deviations) of 50 nm or less.
- The contact micrometer serves as a primary standard for generating commercially available Standard Reference Materials.
- A potential systematic error of 20 nm was identified for the scanning confocal microscope.
Conclusions:
- The developed instruments provide highly accurate and reliable methods for optical fiber cladding diameter measurement.
- The availability of Standard Reference Materials generated by the micrometer facilitates inter-laboratory comparisons and calibration.
- Further investigation into the systematic error of the confocal microscope is warranted for enhanced metrological traceability.
Related Concept Videos
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
369
Assessing safety in wind-exposed installations is crucial to preventing potential failures. This example explores the calculation and design adjustments needed to mount a circular disc on a building facade, where wind forces are a primary concern. A 4-meter diameter disc was initially designed as an aesthetic feature facing winds at a velocity of 25 meters per second, with an air density of 1.25 kilograms per cubic meter. Given these conditions, the drag force on the disc was determined using...
369
Imaging Biological Samples with Optical Microscopy
12.0K
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
12.0K