Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
COMPARATIVE EFFECTIVENESS OF AN ADJUSTABLE TRANSFEMORAL PROSTHETIC INTERFACE ACCOMMODATING VOLUME FLUCTUATION: CASE
Jason T Kahle1, Tyler D Klenow2, M Jason Highsmith3
1OP Solutions, Tampa, FL, USA; Prosthetic Design + Research, Tampa, FL, USA.
An adjustable transfemoral prosthetic socket significantly improves function and comfort for amputees by accommodating volume fluctuations, outperforming the standard of care socket.
Area of Science:
- Prosthetics and Orthotics
- Biomechanical Engineering
- Rehabilitation Medicine
Background:
- The socket-limb interface is critical for amputee stability and mobility.
- Volume fluctuations in the residual limb compromise prosthetic socket fit and function.
- Current prosthetic socket technology lacks effective accommodation for volume changes.
Purpose of the Study:
- To compare the effectiveness of an adjustable transfemoral prosthetic socket against the standard of care (SOC) ischial ramus containment socket.
- To evaluate the sockets' ability to accommodate residual limb volume fluctuations.
- To assess improvements in subjective and performance outcomes.
Main Methods:
- A prospective experimental case study design was utilized.
- Repeated measures of subjective and performance outcomes were collected for both socket conditions.
- The study compared a standard of care socket with an adjustable transfemoral prosthetic interface socket.
Main Results:
- The adjustable socket showed significant improvements (19%-37%) in subjective and performance measures compared to SOC in baseline conditions.
- In volume loss conditions, the adjustable socket improved outcomes by 22%-93%.
- The adjustable socket successfully accommodated simulated volume gain, while the SOC socket failed.
Conclusions:
- The adjustable transfemoral prosthetic interface socket demonstrated superior subjective and performance outcomes compared to the standard of care.
- Adjustable sockets offer a potential solution for managing volume fluctuations in transfemoral amputees.
- Further clinical trials are needed to validate these findings and compare functional outcomes related to volume fluctuation.