Related Experiment Video
Updated: Mar 9, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
131I activity quantification of gamma camera planar images
Raquel Barquero1, Hugo P Garcia1, Monica G Incio1
1Medical Physics, Hospital Clínico Universitario de Valladolid, Avda. Ramón y Cajal 3, 47005 Valladolid, Spain.
This study proposes a method to accurately measure radioactive iodine (I-131) uptake in thyroid tissues during treatment. The findings show that imaging sensitivity depends on lesion size and distance, improving patient treatment accuracy.
Area of Science:
- Nuclear Medicine
- Medical Imaging Physics
- Radiopharmaceutical Therapy
Background:
- Accurate quantification of radiopharmaceutical activity in target tissues is crucial for effective therapeutic administration.
- Thyroid conditions like hyperthyroidism and differentiated thyroid cancer are treated with radioactive iodine (I-131).
- Estimating activity in patients using gamma cameras (GC) requires precise sensitivity calibration.
Purpose of the Study:
- To develop and validate a procedure for estimating I-131 activity in thyroid tissues during treatment.
- To investigate the impact of lesion characteristics (size, distance) on gamma camera sensitivity.
- To reduce uncertainty in measuring thyroid uptake activity for individual patients.
Main Methods:
- Utilized a high-energy (HE) collimator gamma camera (GC) to acquire planar images of I-131.
- Calculated mean fluence (G) using MCNPX code and intrinsic efficiency (η₀) from calibration measurements.
- Estimated sensitivity (S) as G × η₀, validated with SIMIND code and experimental measurements.
Main Results:
- Sensitivity values for Siemens e-cam ranged from 7 to 35 cps MBq⁻¹, and for Philips Skylight from 6 to 29 cps MBq⁻¹.
- Sensitivity was found to be strongly dependent on lesion size (r) and distance (d) from the GC.
- Scatter and penetration photons significantly influenced sensitivity, unlike direct photons.
Conclusions:
- A procedure to estimate patient-specific acquisition sensitivity in thyroidal gamma camera planar imaging was established.
- Individual sensitivity values can significantly reduce uncertainty in measuring thyroid uptake activity.
- This method enhances the accuracy of therapeutic I-131 administration for thyroid conditions.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
10:33Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment
Published on: April 25, 2025
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...