Visualizing the Effects of Sputum on Biofilm Development Using a Chambered Coverglass Model
Published on: December 14, 2016
Effect of the endOclear® Device on Biofilm in Endotracheal Tubes
James M Bardes1, Dana Gray2, Alison Wilson1
11 Division of Trauma, Acute Care Surgery and Critical Care, Department of Surgery, West Virginia University , Morgantown, West Virginia.
Background:
Organisms trapped in biofilms cause more than 80% of medical infections. Significant investments are being made to develop methods of removing these biofilms. The endOclear® device is reported to remove biofilm from endotracheal tubes (ETTs) and to decrease pneumonia rates and ventilator time.
Methods:
This was an observational study performed at a university Level 1 trauma center intensive care unit. A series of 40 ETTs were collected at extubation, with half of the patients having been treated daily with the endOclear® device. Biofilms were quantified from a standardized point on the distal ETT. The patients' standard and biofilm cultures were reviewed.
Results:
The mean hours of intubation for the control group was 135 and for the device group 138. This difference was not statistically significant (p = 0.91). Eleven patients in the device group were found to have pneumonia compared with six in the control group (p = 0.34). Ventilator data after device use showed a mean increase of 29.9 cc in tidal volume and a mean decrease in peak pressures of 0.39 cm H2O. Comparisons between biofilm stage or hours of intubation and a diagnosis of pneumonia found no correlation. Only nine of 40 ETTs had congruence between the microbiata of the biofilm and standard bronchoalveolar lavage (BAL) fluid, a divergence rate of 78%.
Conclusions:
Comparison of the endOclear® group and controls demonstrated a trend toward a higher pneumonia rate in the former. Additionally, the device achieved very small, clinically insignificant, changes in ventilator settings, and no difference was seen in the time on the ventilator. Comparisons between biofilm and standard BAL cultures continue to show the biofilm is more diverse than previously thought. In this study, no statistical significance was found between biofilm stage and the pneumonia rate. This study provides additional evidence that there is no correlation between biofilm stage and duration of intubation.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Biofilms
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...