Standardization from a benchtop to a handheld NIR spectrometer using mathematically mixed NIR spectra to determine
Neirivaldo Cavalcante da Silva1, Cláudia Jéssica Cavalcanti2, Fernanda Araújo Honorato2
1Department of Fundamental Chemistry, Federal University of Pernambuco, Brazil.
Analytica Chimica Acta
|January 14, 2017
Summary
This study introduces virtual standards for calibration transfer in spectroscopy, enabling portable instruments to use data from benchtop models. This method saves time and resources by avoiding physical sample transfer for recalibration.
Area of Science:
- Analytical Chemistry
- Spectroscopy
- Chemometrics
Background:
- Field measurements using portable instruments are increasingly important.
- Calibration transfer techniques allow models from benchtop instruments to be used with portable ones, saving time and cost.
- Traditional methods require physical transfer samples, which can be unstable.
Purpose of the Study:
- To evaluate virtual standards for calibration transfer using the reverse standardization (RS) method.
- To standardize spectral data from dissimilar fuel samples (gasoline, biodiesel blends) between a benchtop and a portable NIR spectrometer.
- To avoid issues associated with physical transfer samples like volatilization and composition changes.
Main Methods:
- Virtual standards were created by mathematically mixing spectra of pure solvents.
- Reverse standardization (RS) was applied to transfer calibration models.
- Partial least squares regression (PLS) models were built for fuel quality parameters.
Main Results:
- Standardization using virtual standards yielded RMSEP values comparable to reference method reproducibility for most parameters.
- The RS procedure successfully transferred spectral data between dissimilar spectrometers.
- Deviations were noted for density and biodiesel content when using virtual samples.
Conclusions:
- The use of virtual standards with the RS method is a promising approach for calibration transfer between benchtop and portable NIR spectrometers.
- This technique facilitates the use of robust benchtop models on portable devices for fuel analysis.
- Virtual standards offer a viable alternative to physical transfer samples, simplifying field measurements.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
1.2K
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
1.2K
IR Spectrometers
3.2K
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
3.2K
UV–Vis Spectrometers
4.2K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
4.2K
Spectrophotometry: Introduction
10.8K
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
10.8K
Flame Photometry: Lab
1.1K
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
1.1K
Flame Photometry: Overview
1.7K
Flame photometry, also known as flame emission spectrometry, is a technique used for the qualitative and quantitative analysis of elements present in a sample using a flame as the source of excitation energy. The concept of flame photometry was realized in the early 1860s by Kirchhoff and Bunsen, who discovered that specific elements emit characteristic radiation when excited in flames. The first instrument developed for this purpose was used to measure sodium (Na) in plant ash using a Bunsen...
1.7K