Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Monitoring Noninvasive Ventilation in Patients with Obesity Hypoventilation Syndrome: Comparison between Ventilator
Ramon Fernandez Alvarez1, Claudio Rabec, Gemma Rubinos Cuadrado
1Servicio de Neumología, Hospital Universitario Central de Asturias, Oviedo, Spain.
Background:
Polygraphy (PG) remains the standard method of assessing noninvasive ventilation (NIV) effectiveness. Built-in software (BIS) of recent NIV equipment provides estimates of some ventilator parameters, but their usefulness is unclear.
Objectives:
To assess the reliability of BIS compared with PG in a cohort of obesity hypoventilation syndrome (OHS) patients on NIV.
Methods:
Thirty stable OHS patients on NIV were evaluated in an outpatient setting with simultaneous PG and BIS recordings. The automated apnea-hypopnea event index (EIAUT) provided by Rescan and manual scoring based on available traces obtained from the software (EIBIS) were compared with manual PG scoring (EIPG). Each manual scoring was separately performed by 2 trained operators. Agreement between the 2 operators was assessed using the kappa coefficient. Pearson correlation and Bland-Altman plots were used to evaluate agreement between EIAUT, EIBIS, and EIPG.
Results:
Twenty-six cases were valid for analysis (age ±61 years, 17 men). All patients were ventilated in the spontaneous/timed mode (mean inspiratory positive airway pressure 17 ± 3 cm H2O, mean expiratory positive airway pressure 10 ± 3 cm H2O). Cohen's kappa agreement between the operators was 0.7 for EIBIS and 0.84 for EIPG. EIBIS showed good correlation with EIPG (r2 = 0.79 p < 0.001), better than scoring provided by the automated analysis (r2 = 0.71, p < 0.006 for EIAUT vs. EIPG).
Conclusions:
In stable OHS patients on NIV, unattended home-based monitoring using Rescan is reproducible and reliable to assess quality of ventilation when compared with PG. In addition, manual scoring of events using data obtained with this device is more consistent than software-based automated analysis.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.