Variable Pulse Duration From a New Holmium:YAG Laser: The Effect on Stone Comminution, Fiber Tip Degradation, and
Daniel A Wollin1, Anika Ackerman1, Chen Yang2
1Division of Urologic Surgery, Duke University Medical Center, Durham, NC.
Objective:
To more clearly define the efficiency and potential benefits of variable pulse-width laser technology for ureteroscopic lithotripsy, we performed comparative in vitro evaluations assessing stone comminution, laser fiber tip degradation, and stone retropulsion.
Methods:
All experiments were conducted using a Swiss LaserClast Holmium:YAG laser (Electro Medical Systems, Nyon, Switzerland) with adjustable pulse duration (300 µs-1500 µs). To assess comminution efficiency and fiber tip degradation, a "dusting" model was employed; the laser fiber tip was moved by a 3-dimensional positioning system in a spiral motion across a flat BegoStone surface submerged in water. Comminution efficiency was measured as the loss of stone mass while fiber tip degradation was measured simultaneously. The same laser and fiber were used in a pendulum model to measure stone retropulsion with a high-speed resolution camera.
Results:
In our dusting model, comminution was significantly greater at high energy (2 J/5 Hz). At the high energy setting, comminution was significantly greater with long pulse duration than short pulse, although this difference was not seen at the high frequency setting (1 J/10 Hz). Tip degradation was increased at high energy settings and was even more pronounced with short pulse duration than long pulse. Short pulse duration caused far more retropulsion than the long pulse setting.
Conclusion:
In an in vitro dusting model, a longer laser pulse duration provides effective stone comminution with the advantage of reducing laser fiber tip degradation and stone retropulsion.
Related Concept Videos
Precipitate Formation and Particle Size Control
The obtained precipitate should be either a pure substance of known composition or easily converted to one by a simple process, such as ignition or drying. In addition, the precipitate should be insoluble and easily filterable. In general, filterability...
Factors Affecting Dissolution: Particle Size and Effective Surface Area
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Compendial Testing Models II