Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Design of Interrogation Protocols for Radiation Dose Measurements Using Optically-Stimulated Luminescent Dosimeters
Sara A Abraham1, Kimberlee J Kearfott, Ali H Jawad
1*University of Michigan, Department of Nuclear Engineering and Radiological Sciences, Ann Arbor, MI 48109-2104; †Department of Physics, University of Azad Jammu and Kashmir, Muzaffarabad, Azad Kashmir 13100, Pakistan; ‡College of Physics and Electronic Engineering, Hengyang Normal University, Hengyang, Hunan Province, China.
Abstract:
Optically-stimulated luminescent dosimeters are capable of being interrogated multiple times post-irradiation. Each interrogation removes a fraction of the signal stored within the optically-stimulated luminescent dosimeter. This signal loss must be corrected to avoid systematic errors in estimating the average signal of a series of optically-stimulated luminescent dosimeter interrogations and requires a minimum number of consecutive readings to determine an average signal that is within a desired accuracy of the true signal with a desired statistical confidence. This paper establishes a technical basis for determining the required number of readings for a particular application of these dosimeters when using certain OSL dosimetry systems.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014