Comparison of a Fast 5-Minute Shoulder MRI Protocol With a Standard Shoulder MRI Protocol: A Multiinstitutional
Naveen Subhas1, Alex Benedick2, Nancy A Obuchowski3
11 Imaging Institute, Cleveland Clinic, 9500 Euclid Ave, A21, Cleveland, OH 44195.
AJR. American Journal of Roentgenology
|February 1, 2017
Summary
A rapid 5-minute shoulder MRI protocol using fast spin-echo (FSE) sequences is as effective as standard MRI for diagnosing shoulder injuries. This faster method shows comparable accuracy and interreader agreement for evaluating tendon and labral tears.
Area of Science:
- Radiology
- Medical Imaging
- Orthopedic Diagnostics
Background:
- Standard shoulder MRI protocols can be time-consuming.
- Optimizing MRI scan times is crucial for patient throughput and comfort.
- Advancements in imaging sequences and hardware offer potential for faster examinations.
Purpose of the Study:
- To compare the diagnostic performance of a novel 5-minute shoulder MRI protocol with a standard protocol.
- To evaluate the interchangeability and accuracy of a fast MRI technique for shoulder imaging.
Main Methods:
- Retrospective review of 201 shoulder MRI examinations (151 at 3-T, 50 at 1.5-T).
- Comparison of a 5-minute fast spin-echo (FSE) MRI protocol with parallel imaging against a standard MRI protocol.
- Assessment of interreader agreement using kappa statistics and comparison of major findings frequency.
Main Results:
- Interprotocol reader agreement for the fast MRI protocol was comparable to the standard protocol (mean difference ≤ 1%).
- Kappa values for the fast MRI were similar to those for the standard MRI (0.373-0.645 vs. 0.320-0.726).
- Sensitivities for tendon and labral tears were equivalent or higher with the fast MRI (33-92%) compared to standard MRI, with similar specificities (77-98%).
Conclusions:
- The 5-minute shoulder MRI protocol using multiplanar 2D FSE sequences with parallel imaging is interchangeable with standard MRI.
- This fast protocol demonstrates similar interreader agreement and diagnostic accuracy for evaluating shoulder injuries.
- Fast MRI offers a viable alternative for efficient and effective shoulder assessment.
Related Concept Videos
Magnetic Resonance Imaging
10.1K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
10.1K
Imaging Studies for Cardiovascular System IV: CMRI
453
Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...