Validation of a Standalone Smartphone Application for Measuring Heart Rate Using Imaging Photoplethysmography
1Cardiio, Inc. , Cambridge, Massachusetts.
Background:
The aim of this study was to determine the accuracy of a freely available smartphone application, Cardiio app (Cardiio, Inc., Cambridge, MA), to measure heart rate from the finger or face using imaging photoplethysmography, by comparing against an FDA-cleared pulse oximeter at rest, and after moderate to vigorous exercise.
Methods:
A total of 40 healthy adults participated in this study. Participants engaged in a period of rest, followed by 3 min of moderate to vigorous intensity cycling on a stationary bicycle. Heart rate measurements were obtained from both the finger and face of participants using the Cardiio app at rest, immediately after exercise, 1-2 min after exercise, and 2-3 min after exercise. Concurrent heart rate readings using an FDA-cleared finger pulse oximeter served as the reference measurement.
Results:
There was a very strong agreement between heart rate measurements obtained using the Cardiio app and the pulse oximeter, both at rest (r = 0.99 for finger, r = 0.97 for face) and after exercise (r = 0.99 for finger, r = 0.97 for face). At rest, the accuracy of the Cardiio app was ±1.58 beats per minute (bpm) (or ±2.27%) using the finger mode and ±2.28 bpm (or ±3.17%) for the face mode, compared to the pulse oximeter. After moderate to vigorous exercise, the accuracy of the Cardiio app was ±2.97 bpm (or ±2.79%) using the finger mode and ±5.31 bpm (or ±4.50%) for the face mode, compared to the pulse oximeter.
Conclusion:
The Cardiio app provided accurate heart rate measurements from the finger and face, both at rest and after exercise.
Related Concept Videos
Special considerations while measuring pulse
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of blood pressure in brachial artery(two-step method)