Imaging of pediatric great vessel stents: Computed tomography or magnetic resonance imaging?
A M den Harder1, D Suchá1, R W van Hamersvelt1
1Department of Radiology, Utrecht University Medical Center, Utrecht, The Netherlands.
Insights
This study evaluated CT and MRI for assessing pediatric great vessel stents in a phantom model. Both modalities showed good stent lumen visibility and reliable measurements, though in-vitro limitations exist.
Area of Science:
- Medical Imaging
- Pediatric Cardiology
- Interventional Radiology
Background:
- Complications can arise after great vessel stent implantation in children.
- Regular imaging follow-up is essential for monitoring these pediatric patients.
Purpose of the Study:
- To identify the optimal imaging technique for evaluating stents in pediatric great vessels.
- This research aims to guide post-procedure monitoring strategies.
Main Methods:
- Five types of large vessel stents were tested in an in-vitro phantom.
- CT and MRI (T2-weighted, T1-weighted, balanced 3D) were used to image the stents.
- Two blinded observers assessed stent lumen visibility and measurement reproducibility.
Main Results:
- CT provided a median stent lumen visibility of 88%.
- MRI's T2-weighted sequence yielded 82% visibility.
- Both CT and MRI demonstrated good interobserver reliability and agreement.
Conclusions:
- CT and MRI are effective in visualizing pediatric great vessel stents in vitro.
- Both imaging modalities offer reliable measurements for stent assessment.
- Further in-vivo studies are needed to evaluate performance in clinical conditions like stenosis.
Background:
Complications might occur after great vessel stent implantation in children. Therefore follow-up using imaging is warranted.
Purpose:
To determine the optimal imaging modality for the assessment of stents used to treat great vessel obstructions in children.
Material And Methods:
Five different large vessel stents were evaluated in an in-vitro setting. All stents were expanded to the maximal vendor recommended diameter (20mm; n = 4 or 10mm; n = 1), placed in an anthropomorphic chest phantom and imaged with a 256-slice CT-scanner. MRI images were acquired at 1.5T using a multi-slice T2-weighted turbo spin echo, an RF-spoiled three-dimensional T1-weighted Fast Field Echo and a balanced turbo field echo 3D sequence. Two blinded observers assessed stent lumen visibility (measured diameter/true diameter *100%) in the center and at the outlets of the stent. Reproducibility of diameter measurements was evaluated using the intraclass correlation coefficient for reliability and 95% limits of agreement for agreement analysis.
Results:
Median stent lumen visibility was 88 (IQR 86-90)% with CT for all stents at both the center and outlets. With MRI, the T2-weighted turbo spin echo sequence was preferred which resulted in 82 (78-84%) stent lumen visibility. Interobserver reliability and agreement was good for both CT (ICC 0.997, mean difference -0.51 [-1.07-0.05] mm) and MRI measurements (ICC 0.951, mean difference -0.05 [-2.52 --2.41] mm).
Conclusion:
Good in-stent lumen visibility was achievable in this in-vitro study with both CT and MRI in different great vessel stents. Overall reliability was good with clinical acceptable limits of agreement for both CT and MRI. However, common conditions such as in-stent stenosis and associated aneurysms were not tested in this in-vitro study, limiting the value of the in-vitro study.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography