Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: An in-vitro study
Fayez Elkholy1, Falko Schmidt1, Rudolf Jäger1
1Department of Orthodontics, University of Ulm, Ulm, Germany.
Introduction:
Recent studies have shown that therapeutic loads applied to individual teeth by aligners may substantially exceed recommended values. The primary purpose of this study was to quantify force and moment components during derotation of a maxillary central incisor when 0.3-mm-thick or 0.4-mm-thick polyethylene terephthalate glycol aligners were used instead of conventional polyethylene terephthalate glycol aligners with a minimum thickness of 0.5 mm.
Methods:
The test setup consisted of an acrylic model of a maxilla with a separated right central incisor mounted on a 3-dimensional force and moment sensor. The force and moment components were recorded for aligners with thicknesses ranging from 0.3 to 0.75 mm during ±10° rotation and derotation of the separated incisor.
Results:
Moments exerted by the thinnest aligner currently available, 0.5 mm, were 73.57 Nmm for the 10° mesiorotation. In comparison, the corresponding moments with the 0.4-mm and 0.3-mm aligners were 41.08 and 17.84 Nmm, respectively. Moment values for derotation of the maxillary right central incisor into neutral position showed nonlinear return curves indicating viscoelastic material behavior.
Conclusions:
A significant load reduction can be achieved with the new thinner aligners. Because of the form instability of the 0.3-mm aligner during handling, we suggest the novel sequence 0.4, 0.5, and 0.75 mm for aligner systems based on sequentially increased material thickness. This sequence combines sufficiently low initial aligner stiffness and steady load increases in single setup steps. The viscoelastic behavior of polyethylene terephthalate glycol aligners observed during incisor derotation should lead to a reduction of the high initial load exerted directly after intraoral aligner insertion.
Related Concept Videos
Resultant Moment: Scalar Formulation
To determine the resultant moment, the moments caused by all the forces in a system in the x-y plane are considered. Positive moments are typically...
Deformation in a Circular Shaft
Plastic Deformations of Members with a Single Plane of Symmetry
Bearings: Problem Solving
Eccentric Axial Loading in a Plane of Symmetry
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...