Correlation between serum metal ion levels and adverse local tissue reactions after Conserve® Plus hip resurfacing
Harlan C Amstutz1, Michel J Le Duff1
1Joint Replacement Institute at St. Vincent Medical Center, Los Angeles - USA.
Background:
Adverse local tissue reactions (ALTR) have been associated with the use of metal-on-metal (MoM) bearings and the monitoring of cobalt (Co) and chromium (Cr) ion levels in blood or serum may be the best way to evaluate in vivo the wear of these bearings. However, the relationship between Co and Cr ion concentrations and the formation of ALTR remains unclear.
Methods:
We investigated the relationship between ALTR and serum Co and Cr ion levels and identified the clinical factors influencing the formation of ALTR in patients treated with MoM hip resurfacing arthroplasties. 228 patients with unilateral Conserve® Plus MoM hip resurfacing had serum metal ion studies performed more than 1 year after surgery. Metal artifact reduction sequence magnetic resonance imaging (MARS MRI) was performed on subjects at risk for ALTR as determined by a screening protocol.
Results:
12 patients had ALTR. Logistic regression showed a strong association of ALTR with elevated ion levels and with low (<10 mm) contact patch to rim distance.
Conclusions:
MoM bearings require enough functional coverage of the socket by design and then precise implantation to maximise functional coverage of the femoral ball, enhance lubrication, and avoid edge-loading wear.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Related Concept Videos
EDTA: Auxiliary Complexing Reagents
Extraction: Advanced Methods