Carbon dioxide washout during high flow nasal cannula versus nasal CPAP support: An in vitro study
Emidio M Sivieri1,2, Elizabeth E Foglia2,3, Soraya Abbasi1,2,3
1Section on Newborn Pediatrics, Pennsylvania Hospital, Philadelphia, Pennsylvania.
Objective:
To compare CO2 washout time at different levels of HFNC versus NCPAP in a premature infant lung model with simulated mouth-closed and mouth-open conditions using two sizes of nasal cannula and full- and half-prong HFNC insertion depths.
Design/Methods:
A piston-cylinder lung simulator, having a fixed volume of 30 ml and a 4.8 ml dead space, simulated spontaneous breathing (6.5 ml tidal volume, 50 br/min, Ti = 0.5 sec). Two Fisher & Paykel™ cannulas (Fisher & Paykel Healthcare Ltd., Auckland, New Zealand) (2.8 and 3.2 mm O.D.) and two Infant-Flow™ (CareFusion, Yorba Linda, CA) NCPAP cannulas (3.4 and 4.1 mm O.D.) were applied to simulated airways having either 3.5 or 4.5 mm I.D. nares. Simulated mouth opening was a 5 mm I.D. side tap below the nasal interface. The lung was primed with 5% CO2 . Washout times were determined at HFNC settings of 3, 4, 5, 6, and 8 L/min and NCPAP at 3, 4, 5, 6, and 8 cm H2 O with simulated open and closed-mouth conditions and full- and half-inserted HFNC prongs.
Results:
Overall combined mean washout times for NCPAP with mouth-closed were significantly longer than HFNC over all five pressure and flow device settings by 16.2% (P < 0.001). CO2 washout times decreased as flow or pressure device settings were increased. There were negligible differences in washout times between NCPAP and HFNC with mouth-open. Mouth-open washout times were significantly less than mouth-closed for all conditions. Overall closed-mouth washout times for HFNC half-prong insertion were longer than for full-prong insertion by 5.3% (P < 0.022).
Conclusions:
Significantly improved CO2 elimination using HFNC versus NCPAP should be a particularly important consideration in premature infants having very high dead space-to-tidal volume ratio compared to larger infants. Pediatr Pulmonol. 2017;52:792-798. © 2017 Wiley Periodicals, Inc.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...