Related Experiment Video
Updated: Mar 7, 2026

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Initial Experience of a Second-Generation Self-Expanding Transcatheter Aortic Valve: The UK & Ireland Evolut R
Sundeep S Kalra1, Sami Firoozi1, James Yeh2
1Cardiology Clinical Academic Group, St. George's University Hospitals NHS Foundation Trust & St. George's University of London, London, United Kingdom.
Objectives:
The authors present the UK and Irish real-world learning curve experience of the Evolut R transcatheter heart valve.
Background:
The Evolut R is a self-expanding, repositionable, and fully recapturable second-generation transcatheter heart valve with several novel design features to improve outcomes and reduce complications.
Methods:
Clinical, procedural, and 30-day outcome data were prospectively collected for the first 264 patients to receive the Evolut R valve in the United Kingdom and Ireland.
Results:
A total of 264 consecutive Evolut R implantations were performed across 9 centers. The mean age was 81.1 ± 7.8 years, and the mean logistic European System for Cardiac Operative Risk Evaluation score was 19.9 ± 13.7%. Procedural indications included aortic stenosis (72.0%), mixed aortic valve disease (17.4%), and failing aortic valve bioprostheses (10.6%). Conscious sedation was used in 39.8% of patients and transfemoral access in 93.6%. The procedural success rate was 91.3%, and paravalvular leak immediately after implantation was mild or less in 92.3%. Major complications were rare: cardiac tamponade in 0.4%, conversion to sternotomy in 0.8%, annular rupture in 0.0%, coronary occlusion in 0.8%, major vascular in 5.3%, acute kidney injury in 6.1%, new permanent pacemaker implantation in 14.7%, and procedure-related death in 0.0%. At 30-day follow-up, survival was 97.7%, paravalvular leak was mild or less in 92.3%, and the stroke rate was 3.8%.
Conclusions:
This registry represents the largest published real-world experience of the Evolut R valve. The procedural success rate was high and safety was excellent, comparable with previous studies of the Evolut R valve and other second-generation devices. The low rate of complications represents an improvement on first-generation devices.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Aortic Regurgitation I: Introduction
Cardiac Catheterization I: Pre-Procedure Overview