Technical Note: Fully-automated analysis of Jaszczak phantom measurements as part of routine SPECT quality control
Albert Hirtl1, Helmar Bergmann2, Barbara Knäusl3
1Division of Nuclear Medicine, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Austria.
Purpose:
Inspection and quantitative validation of tomographic imaging properties of SPECT systems, i.e., spatial resolution, contrast, and inhomogeneity must be performed in regular intervals. Typically, the modular Jaszczak phantom is used for that purpose, as it offers the possibility to investigate all three system properties with a single measurement. The interpretation of the measurement is performed visually, thus, being insensitive to subtle changes in system performance. To overcome this limitation, a fully-automated software for the objective analysis of Jaszczak phantom measurements is proposed here.
Methods:
The software was developed as an ImageJ plugin and offers a number of sequential evaluation steps: automatic determination of the type of Jaszczak phantom, calculation of sector and sphere contrast, detection of ring artifacts using either the Hough transform, followed by a threshold-based decision criterion, or Student's t-test. Monte Carlo simulations were used to estimate the detectability limits for ring artifacts.
Results:
The software successfully calculated sector and sphere contrasts and reliably determined ring artifacts present in the homogeneity part of the Jaszczak phantom, based on automatic identification of the phantom type.
Conclusion:
Given the quantitative nature of the produced output, results from one imaging system can easily be compared to another in an objective way. The advantage of the software is clearly that the information provided is objective and does not rely on the experience level of the user.
More Related Videos
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight MALDI-TOF Mass Spectrometry
Published on: June 10, 2018
07:55Author Spotlight: A Tailor-Made Sample Preparation Approach for Enhanced MALDI-IMS Analysis of Hard Palm Seeds
Published on: June 30, 2023
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Mass Spectrometers
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...