Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
Ambient Pressure Inverse Ion Mobility Spectrometry Coupled to Mass Spectrometry
Wenjie Liu1,2, Austen L Davis1, William F Siems1
1Department of Chemistry, Washington State University , Pullman, Washington 99163, United States.
Abstract:
Although higher resolving powers are often achieved using ambient pressure drift tube ion mobility mass spectrometry (DT-IMMS) systems, lower duty cycles are often required which directly impacts sensitivity. Moreover, the mechanism of ion gating using Bradbury-Nielsen or Tyndall-Gate configurations routinely results in ion gate depletion effects which discriminate against low mobility ions. This paper reports a new method of ambient pressure ion mobility operation in which inverse ion mobility spectrometry is coupled to a time-of-flight mass spectrometer to improve sensitivity and minimize the effects of ion gate depletion. In this mode of operation, the duty cycle is improved to approximate 99% from a typical value of less than 1%, improving the signal intensity by over 2 orders of magnitude. Another advantage of inverse ion mobility mass spectrometry is a reduction of the impact of ion gate depletion on low mobility molecules that translates into higher sensitivity for this class of analytes. To demonstrate these benefits afforded by this instrumental mode of operation differences in sensitivity, resolving power, and ion discrimination are compared between the inverse and normal modes of operation using tetraalkylammonium standards. These results show that the ion throughput is significantly increased for analytes with a broad range of mobilities with little impact on resolving power. While the mobility-based discrimination is minimized using the inverse mode of operation, the noise level in the inverse mode is highly dependent upon the stability of ionization source.
More Related Videos
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Overview
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometers
Chemical Ionization (CI) Mass Spectrometry