A Novel Apparatus for the Multifaceted Evaluation of Arterial Function Through Transmural Pressure Manipulation
Toshiyuki Yaguchi1,2, Yalin Cong3, Kenji Shimo3
1Biomechanics Laboratory, Department of Mechanical Engineering, Nagoya Institute of Technology, Nagoya, 466-8555, Japan. tyaguchi@mail.dendai.ac.jp.
Abstract:
A novel apparatus for the multifaceted evaluation of artery function was developed. It measures endothelial and smooth muscle functions and the pressure-strain elastic modulus (E p). A rigid airtight chamber with an ultrasound probe was attached to the upper arm to manipulate the transmural pressure of the brachial artery. Endothelial function was measured via a standard flow-mediated dilation (FMD) protocol. Smooth muscle function was evaluated via a myogenic contraction of the artery following the application of negative pressure to the chamber and was named pressure-mediated contraction (PMC). E p was obtained by measuring the instantaneous increase in the artery diameter following the negative pressure application. The PMC and FMD values had a significant negative correlation with age, indicating that the age-related decrease in FMD is caused by the decay of endothelial and smooth muscle function. A consideration of PMC may help improve the accuracy of artery function measurement. E p in subjects aged >40 years was found to be significantly higher in the supra-physiological pressure range than in the physiological one (p = 0.02); this did not occur in younger subjects. Artery stiffening may begin in the supra-physiological range, and this stiffness may also be used for the diagnosis of atherosclerosis.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: