Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Water and lipid diffusion MRI using chemical shift displacement-based separation of lipid tissue (SPLIT)
Naoki Ohno1, Hirohito Kan2, Tosiaki Miyati1
1Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 9200942, Japan.
Purpose:
To obtain water and lipid diffusion-weighted images (DWIs) simultaneously, we devised a novel method utilizing chemical shift displacement-based separation of lipid tissue (SPLIT) imaging.
Materials And Methods:
Single-shot diffusion echo-planar imaging without fat suppression was used and the imaging parameters were optimized to separate water and lipid DWIs by chemical shift displacement of the lipid signals along the phase-encoding direction. Using the optimized conditions, transverse DWIs at the maximum diameter of the right calf were scanned with multiple b-values in five healthy subjects. Then, apparent diffusion coefficients (ADCs) were calculated in the tibialis anterior muscle (TA), tibialis bone marrow (TB), and subcutaneous fat (SF), as well as restricted and perfusion-related diffusion coefficients (D and D*, respectively) and the fraction of the perfusion-related diffusion component (F) for TA.
Results:
Water and lipid DWIs were separated adequately. The mean ADCs of the TA, TB, and SF were 1.56±0.03mm2/s, 0.01±0.01mm2/s, and 0.06±0.02mm2/s, respectively. The mean D*, D, and F of the TA were 13.7±4.3mm2/s, 1.48±0.05mm2/s, and 4.3±1.6%, respectively.
Conclusion:
SPLIT imaging makes it possible to simply and simultaneously obtain water and lipid DWIs without special pulse sequence and increases the amount of diffusion information of water and lipid tissue.