The Measurement of X-ray Fluoroscopy Input-output Characteristics in Angiography System
Yoshiaki Hagihara1, Takayoshi Umehara, Kazuma Matsumoto
1Department of Radiological Technology, Hyogo College of Medicine College Hospital.
Purpose:
Determination of X-ray fluoroscopy radiation dose and contrast with angiographic system automatically depending on the objects, and to control setting manually, which is difficult for the measurement of characteristics. Therefore, we examined the method to adjust the conditions of fluoroscopy and measured the input-output characteristics.
Method:
To adjust and fix the condition of fluoroscopy, the exposure adjustment area at the center of the irradiation field was moved to the left side and attached the copper plates to regulate the exposure dose. The area to measure the digital value was selected at the center of the irradiation field, and the dosimeter was placed at the right side of the area, which was selected to measure the digital value. To regulate the entrance dose progressively, the acryl plates were inserted into the irradiation field except for the exposure adjustment area. We obtained a characteristic curve from the measured dose and the digital value. Difference of lookup table (LUT), dose dependency, and tube voltage dependency were checked by the digital characteristic curves.
Result:
Each LUT showed different curves, but they all saturated with 4095, which is the maximum value of 12 bits. Dose dependency was measured as an increase in the permitted dose level with an increase in the setting dose. Tube voltage dependency improved with the tube voltage rises. Each characteristic curve became same by converting the relative exposure dose. As a result, measuring the shape of LUT would be possible.
Conclusion:
The method is useful for measuring the characteristic curve with the X-ray fluoroscopy of angiographic system.
More Related Videos
05:26Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
Published on: January 7, 2019
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography