Related Experiment Video
Updated: Mar 7, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
An Analytical Method to Calculate Phantom Scatter Factor for Photon Beam Accelerators
Mohammad Javad Tahmasebi Birgani1, Nahid Chegeni2, Mohammad Ali Behrooz3
1Ph.D. of Medical Physics, Professor, Department of Medical Physics and Radiation Therapy and Medical Physics Department, Golestan Hospital, Jundishapur University of Medical Sciences, Faculty of Medicine, Ahvaz Jundishapur University of Medical Sciences: Ahvaz, Iran.
Introduction:
One of the important input factors in the commissioning of the radiotherapy treatment planning systems is the phantom scatter factor (Sp) which requires the same collimator opening for all radiation fields. In this study, we have proposed an analytical method to overcome this issue.
Methods:
The measurements were performed using Siemens Primus Plus with photon energy 6 MV for field sizes from 5×5cm2 to 40×40cm2. Phantom scatter factor was measured through the division of total scatter output factors (Scp), and collimator scatter factor (Sc).
Results:
The mean percent difference between the measured and calculated Sp was 1.00% and -3.11% for 5×5, 40×40 cm2 field size respectively.
Conclusion:
This method is applicable especially for small fields used in IMRT which, measuring collimator scatter factor is not reliable due to the lateral electron disequilibrium.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
11:34Controlled Synthesis and Fluorescence Tracking of Highly Uniform PolyN-isopropylacrylamide Microgels
Published on: September 8, 2016