Related Experiment Video
Updated: Mar 7, 2026

Author Spotlight: Development of a Novel Finite Element Analysis Model for Improved Orthognathic Surgical Techniques
Published on: October 20, 2023
The CLS uncemented expansion cup: six cases of component failure
1Hip Surgery Unit, Department of Orthopaedics and Trauma, Kingston Hospital, Surrey UK.
Abstract:
We report six cases of fracture of the metal backing of the CLS expansion cup. The aim of this paper is to draw attention to this mode of failure and recommend that in cases of unexplained hip pain in those with a CLS acetabular component X-rays in at least two planes are performed. (Hip International 2004; 14: 28-33).