Development of high throughput LC/MS/MS method for analysis of perfluorooctanoic acid from serum, suitable for
Eduard Rogatsky1, Colleen O'Hehir1, James Daly1
1Wadsworth Center, New York State Department of Health, Division of Environmental Health Sciences, Albany, NY 12201, United States.
Abstract:
A simple method for determination of perfluorooctanoic acid (PFOA) from human serum by liquid chromatography/tandem mass spectrometry (LC/MS/MS) and utilizing high-throughput sample preparation was developed by New York State Public Health Emergency Preparedness Laboratory and used for biomonitoring studies. 50μl of serum was mixed with a 0.4ml of acetonitrile containing isotopically labeled internal standard, followed by phospholipid removal/protein precipitation. The extract was partially dried and analyzed by LC/MS/MS. The linear range of PFOA analysis was 0.5-100ng/ml. LLOQ was selected as 0.5ng/ml. The method was validated following APHL guidelines for LRN-C laboratories and about 6000 specimens were successfully prepared and analyzed using this simple LC/MS method. Due to sample preparation simplicity, the presented method can be used in large-scale clinical testing, such as public health surveillance studies.