Cost-Effectiveness Analysis of Quadripolar Versus Bipolar Left Ventricular Leads for Cardiac Resynchronization
Jonathan M Behar1, Hui Men Selina Chin2, Steve Fearn3
1Imaging Sciences & Biomedical Engineering, King's College London, London, United Kingdom; Department of Cardiology, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
Objectives:
The objective of this study was to evaluate the cost-effectiveness of quadripolar versus bipolar cardiac resynchronization defibrillator therapy systems.
Background:
Quadripolar left ventricular (LV) leads for cardiac resynchronization therapy reduce phrenic nerve stimulation (PNS) and are associated with reduced mortality compared with bipolar leads.
Methods:
A total of 606 patients received implants at 3 UK centers (319 Q, 287 B), between 2009 and 2014; mean follow-up was 879 days. Rehospitalization episodes were costed at National Health Service national tariff rates, and EQ-5D utility values were applied to heart failure admissions, acute coronary syndrome events, and mortality data, which were used to estimate quality-adjusted life-year differences over 5 years.
Results:
Groups were matched with regard to age and sex. Patients with quadripolar implants had a lower rate of hospitalization than those with bipolar implants (42.6% vs. 55.4%; p = 0.002). This was primarily driven by fewer hospital readmissions for heart failure (51 [16%] vs. 75 [26.1%], respectively, for quadripolar vs. bipolar implants; p = 0.003) and generator replacements (9 [2.8%] vs. 19 [6.6%], respectively; p = 0.03). Hospitalization for suspected acute coronary syndrome, arrhythmia, device explantation, and lead revisions were similar. This lower health-care utilization cost translated into a cumulative 5-year cost saving for patients with quadripolar systems where the acquisition cost was <£932 (US $1,398) compared with bipolar systems. Probabilistic sensitivity analysis results mirrored the deterministic calculations. For the average additional price of £1,200 (US $1,800) over a bipolar system, the incremental cost-effective ratio was £3,692 per quality-adjusted life-year gained (US $5,538), far below the usual willingness-to-pay threshold of £20,000 (US $30,000).
Conclusions:
In a UK health-care 5-year time horizon, the additional purchase price of quadripolar cardiac resynchronization defibrillator therapy systems is largely offset by lower subsequent event costs up to 5 years after implantation, which makes this technology highly cost-effective compared with bipolar systems.