Effectiveness of Pulse Oximetry Versus Doppler for Tourniquet Monitoring
Background:
Pulse oximeters are common and include arterial pulse detection as part of their methodology. The authors investigated the possible usefulness of pulse oximeters for monitoring extremity tourniquet arterial occlusion.
Methods:
Tactical Ratcheting Medical Tourniquets were tightened to the least Doppler-determined occluding pressure at mid-thigh or mid-arm locations on one limb at a time on all four limbs of 15 volunteers. A randomized block design was used to determine the placement locations of three pulse oximeter sensors on the relevant digits. The times and pressures of pulsatile signal absences and returns were recorded for 200 seconds, with the tourniquet being tightened only when the Doppler ultrasound and all three pulse oximeters had pulsatile signals present (pulsatile waveform traces for the pulse oximeters).
Results:
From the first Doppler signal absence to tourniquet release, toe-located pulse oximeters missed Doppler signal presence 41% to 50% of the times (discrete 1-second intervals) and missed 39% to 49% of the pressure points (discrete 1mmHg intervals); fingerlocated pulse oximeters had miss rates of 11% to 15% of the times and 13% to 19% of the pressure points. On toes, the pulse oximeter ranges of sensitivity and specificity for Doppler pulse detection were 71% to 90% and 44% to 51%, and on fingers, the respective ranges were 65% to 77% and 78% to 83%.
Conclusion:
Use of a pulse oximeter to monitor limb tourniquet effectiveness will result in some instances of an undetected weak arterial pulse being present. If a pulse oximeter waveform is obtained from a location distal to a tourniquet, the tourniquet should be tightened. If a pulsatile waveform is not detected, vigilance should be maintained.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse