Comparison Between Wrist-Worn and Waist-Worn Accelerometry
Related Experiment Video
Updated: Mar 6, 2026

A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Objective:
To use the most recent ActiGraph model (GT9X) to compare counts per minute (CPM) estimates between wrist-worn and waist-worn attachment sites.
Methods:
Participants completed 2 conditions (laboratory [N = 13] and free-living conditions [N = 9]), in which during both of these conditions they wore 2 ActiGraph GT9X accelerometers on their nondominant wrist (side-by-side) and 2 ActiGraph GT9X accelerometers on their right hip in line with the midaxillary line (side-by-side). During the laboratory visit, participants completed 5 treadmill-based trials all lasting 5 min: walk at 3 mph, 3.5 mph, 4 mph, and a jog at 6 mph and 6.5 mph. During the free-living setting, participants wore the monitors for 8 hours. Paired t test, Pearson correlation and Bland-Altman analyses were employed to evaluate agreement of CPM between the attachment sites.
Results:
Across all intensity levels and setting (laboratory and free-living), CPM were statistically significantly and substantively different between waist- and wrist-mounted accelerometry.
Conclusion:
Attachment site drastically influences CPM. As such, extreme caution should be exercised when comparing CPM estimates among studies employing different attachment site methodologies, particularly waist versus wrist.
Related Concept Videos
Measuring Acceleration Due to Gravity
A simple pendulum can be described as a point mass and a string. Meanwhile, a physical pendulum is any object whose oscillations are similar to a simple pendulum, but cannot be modeled as a point mass on a string because its mass is distributed over a larger area. The behavior of a physical pendulum can be modeled using the principles of...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Relative Motion Analysis - Acceleration