Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Diffusion-prepared stimulated-echo turbo spin echo (DPsti-TSE): An eddy current-insensitive sequence for
Qinwei Zhang1, Bram F Coolen2, Maarten J Versluis3
1Department of Radiology, Academic Medical Center, Amsterdam, the Netherlands.
A new 3D diffusion-prepared stimulated-echo (DPsti-TSE) sequence offers high-resolution, undistorted diffusion-weighted imaging (DWI). This technique minimizes eddy current effects, providing accurate apparent diffusion coefficient (ADC) values and a promising alternative to traditional methods.
Area of Science:
- Magnetic Resonance Imaging
- Diffusion-Weighted Imaging
Background:
- Traditional 2D diffusion-weighted echo planar imaging (DW-EPI) suffers from signal loss due to eddy currents, limiting image quality and accuracy.
- Developing advanced MRI sequences is crucial for improving diagnostic capabilities in various clinical applications.
Purpose of the Study:
- To introduce and validate a novel three-dimensional (3D), diffusion-prepared turbo spin echo sequence with a stimulated-echo readout (DPsti-TSE).
- To demonstrate the capability of DPsti-TSE for high-resolution and undistorted diffusion-weighted imaging (DWI), overcoming limitations of conventional methods.
Main Methods:
- Implementation of a DPsti-TSE sequence incorporating a dephasing gradient in the diffusion preparation and rephasing gradients in the turbo spin echo module.
- Validation using phantom experiments to assess apparent diffusion coefficient (ADC) values and transient signal behavior.
- In vivo imaging of prostate and carotid vessel wall to evaluate resolution, distortion, and scan time.
Main Results:
- DPsti-TSE demonstrated near-perfect agreement with DW-EPI for ADC values in phantom studies.
- High-resolution, undistorted 3D whole-prostate DWI was achieved in 6 minutes with four b values.
- Consistent ADC values were obtained for carotid vessel wall imaging in 12 minutes, showing robustness across subjects and slice locations.
Conclusions:
- DPsti-TSE is a robust 3D diffusion-weighted sequence that effectively minimizes eddy current-induced signal loss.
- The sequence provides high-resolution, undistorted DWI with accurate ADC quantification.
- DPsti-TSE presents an attractive alternative to conventional 2D DW-EPI for various clinical applications, including prostate and vascular imaging.