Related Experiment Video
Updated: Mar 6, 2026

Author Spotlight: Advanced Techniques for Characterizing Tissue Mineralization in Bone Regeneration Research
Published on: September 27, 2024
Feasibility of assessing bone matrix and mineral properties in vivo by combined solid-state 1H and 31P MRI
Xia Zhao1, Hee Kwon Song1, Alan C Seifert1
1Laboratory for Structural, Physiologic and Functional Imaging, Department of Radiology, Perelman School of Medicine, University of Pennsylvania, 1 Founders Building, MRI Education Center, Philadelphia, PA, United States of America.
Purpose:
To develop and evaluate an integrated imaging protocol for bone water and phosphorus quantification in vivo by solid-state 1H and 31P MRI.
Materials And Methods:
All studies were HIPAA-compliant and were performed with institutional review board approval and written informed consent. Proton (1H) ultra-short echo-time (UTE) and phosphorus (31P) zero echo-time (ZTE) sequences were designed and implemented on a 3 T clinical MR scanner to quantify bone water and mineral in vivo. The left tibia of ten healthy subjects (including both genders, 49±15 y/o) was examined with a custom-built 1H/31P dual-frequency extremity RF coil. Total bone water (TW), water bound to the collagen matrix (BW) and bone 31P were quantified from MR images with respect to reference samples of known 1H or 31P concentration, and pore water (PW) was subsequently determined from TW and BW. Porosity index (PI) was calculated as the ratio between UTE images acquired at two echo times. MRI parameters were compared with bone density measures obtained by high-resolution peripheral quantitative CT (HR-pQCT).
Results:
The total scan time for the bone water and 31P quantification protocol was about 50 minutes. Average TW, BW, PW and 31P concentrations were 13.99±1.26, 10.39±0.80, 3.34±1.41 mol/L and 7.06±1.53 mol/L for the studied cohort, respectively, in good agreement with previous results conducted ex vivo. Average intra-subject coefficients of variation were 3.47%, 2.60% and 7.50% for TW, BW and PW and 5.60% for 31P. Negative correlations were observed between PW and vBMD (p<0.05) as well as between PI and 31P (p<0.05), while bone mineral content (BMC) estimated from 31P MRI and HR-pQCT were strongly positively correlated (p<0.0001).
Conclusion:
This work demonstrates the feasibility of quantifying bone water and mineral phosphorus in human subjects in a single MRI session with a clinically practical imaging protocol.