Analytical comparison between two hematological analyzer systems: CAL-8000 vs. XN-9000
S Buoro1, T Mecca1, M Seghezzi1
1Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
International Journal of Laboratory Hematology
|March 16, 2017
Summary
This study compared CAL-8000 and XN-9000 hematology analyzers, finding significant differences in morphology flag generation. Harmonization or instrument-specific ranges are needed for accurate automated differential leukocyte counts.
Area of Science:
- Hematology
- Clinical Chemistry
- Medical Laboratory Science
Background:
- Automated hematology analyzers are crucial for differential leukocyte counts (DIFF profile).
- Morphological flags aid in identifying abnormal cells, but their performance varies between instruments.
- Comparing new (CAL-8000) and established (XN-9000) analyzers is essential for diagnostic accuracy.
Purpose of the Study:
- To compare the analytical performance of CAL-8000 and XN-9000 hematology analyzers.
- To evaluate traditional and new parameters, including morphological flags, against optical microscopy (OM).
- To assess the agreement of automated DIFF profiles with manual microscopic review.
Main Methods:
- 1025 peripheral blood samples were analyzed on CAL-8000, XN-9000, and by OM.
- Within-run imprecision was assessed in low cellularity samples.
- Statistical analyses included Spearman's correlation, Passing-Bablok regression, Bland-Altman bias, and Cohen's K test.
Main Results:
- CAL-8000 and XN-9000 showed reproducible data for within-run imprecision, except for low platelet counts.
- Passing-Bablok regression revealed varying slopes and intercepts between analyzers for leukocyte, RBC, platelet, and reticulocyte parameters.
- Morphological flag comparison yielded Cohen's K values <0.55, indicating poor agreement; DIFF profiles showed bias ranging from -42.9% to 35.0% compared to OM.
Conclusions:
- Significant analytical differences exist between the CAL-8000 and XN-9000 analyzers.
- The generation of morphology flags showed particular discordance between the instruments.
- Harmonization of analytical methods or adoption of instrument-specific reference ranges is recommended for improved diagnostic consistency.