Impaired clinical utility of sequential patient GEM blood gas measurements associated with calibration schedule
George S Cembrowski1, Qian Xu2, Adam R Cembrowski3
1Department of Laboratory Medicine and Pathology, University of Alberta Hospital, 4B1.24 Walter C. Mackenzie Centre, 8440 - 112 Street, Edmonton, Alberta T6G 2B7, Canada; University of Alberta, 4B1.24 Walter C. Mackenzie Centre, 8440 - 112 Street, Edmonton, Alberta T6G 2B7, Canada; Alberta Health Services, 4B1.24 Walter C. Mackenzie Centre, 8440 - 112 Street, Edmonton, Alberta T6G 2B7, Canada.
Background:
Within- and/or between-instrument variation may falsely indicate patient trends or obscure real trends. We employ a methodology that transforms sequential intra-patient results into estimates of biologic and analytic variation. We previously derived realistic biologic variation (sb) of blood gas (BG) and hematology analytes. We extend this methodology to derive the imprecision of two GEM 4000 BG analyzers.
Methods:
A laboratory data repository provided arterial BG, electrolyte and metabolite results generated by two GEM 4000s on ICU patients in 2012-2013. We tabulated consecutive pairs of intra-patient results separated by increasing time interval between consecutive tests. The average between pair variations were regressed against time with the y-intercept representing the sum of the biologic variation and short term analytic variation: yo2=sb2+sa2. Using an equivalent equation for the Radiometer ABL, the imprecision of the two GEMs was calculated: saGEM=(yoGEM2-yoABL2+saABL2)1/2. This analysis was performed for nearly all measurements, regardless of time as well for values obtained over two 12h mutually exclusive periods, starting either at 2am or 2pm.
Results:
Regression graphs were derived from 1800 patients' blood gas results with least 10,000 data pairs grouped into 2h intervals. The calculated saGEM exceed the directly measured saABL with many GEM sigma ratios of biologic variation/analytic variation being close to unity. All of the afternoon saGEM exceeded their morning counterparts with pH, pCO2, K and bicarbonate being statistically significant.
Conclusion:
For many analytes, the average analytical variation of tandem GEMs approximates the biologic variation, indicating impaired clinical usefulness of tandem sequential measurements. A significant component of this variation is due to increased variation of the GEMs between 2pm and 2am.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.