Calibration procedure for secular frequency scanning in ion trap mass spectrometers
Dalton T Snyder1, Christopher J Pulliam1, R Graham Cooks1
1Department of Chemistry and Center for Analytical Instrumentation Development, Purdue University, West Lafayette, IN, 47907, USA.
A new algorithm simplifies mass calibration for ion trap mass spectrometry using secular frequency scans. This method improves accuracy and addresses challenges in mass assignment for trapped ions.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
Background:
- Ion traps can record mass spectra by scanning alternating current (ac) signal frequencies, simplifying instrumentation compared to radiofrequency (rf) amplitude scans.
- Mass calibration presents a significant challenge in this technique.
Purpose of the Study:
- To investigate an algorithm for mass calibration in ion trap mass spectrometry using secular frequency scans.
- To discuss factors influencing ion mass assignments.
Main Methods:
- Time domain data from a linear ion trap mass spectrometer were converted to the m/z domain using ion Mathieu parameter q_u values.
- An algorithm derived from Mathieu equations was developed to determine q_u, m/z values, and peak widths through successive approximations.
- Experiments involved sweeping an auxiliary ac signal frequency to induce mass-selective ion ejection.
Main Results:
- The developed algorithm achieved calibration accuracy better than 0.1%.
- Mass resolution ranged from 250 to 500 (m/z 100-1800) in secular frequency scans.
- The algorithm's predictions were validated against experimental data.
Conclusions:
- A simple and effective mass calibration method for secular frequency scans in ion traps was successfully implemented.
- The study analyzed the impact of rf amplitude, scan rate, and ac amplitude on calibration.
- Accurate calibration requires corrections for ion mass differences, which are easily integrated into the procedure.