Alternative mass reference standards for direct analysis in real time mass spectrometry
1JEOL USA, Inc., 11 Dearborn Rd., Peabody, MA, 01960, USA.
Direct Analysis in Real Time (DART) mass spectrometry calibration is improved using Jeffamine-M600 for positive ions and Fomblin Y for negative ions. These standards offer convenient and effective mass calibration across various m/z ranges.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
Background:
- Mass spectrometry requires reliable calibration standards for accurate mass measurements.
- Direct Analysis in Real Time (DART) is a versatile ionization technique.
- Existing standards may have limitations in specific mass ranges or ionization modes.
Purpose of the Study:
- To evaluate amine-terminated polyether and fluorinated materials as mass calibration standards for DART mass spectrometry.
- To identify effective positive-ion and negative-ion reference standards for DART analysis.
Main Methods:
- Mass spectra were acquired using a time-of-flight mass spectrometer with a DART ion source.
- Reference standards were analyzed by direct insertion into the DART gas stream.
- Commercial materials including Jeffamine® M-600, PFTBA, Ultramark 1621®, and Fomblin® Y were tested.
Main Results:
- Jeffamine® M-600 produced intense peaks for protonated molecules, suitable for positive-ion calibration.
- Fomblin® Y provided useful negative-ion reference spectra from m/z 85 to >3000.
- DART mass spectra of Ultramark 1621® were consistent with previous FAB and ESI results.
Conclusions:
- Jeffamine-M600 is a convenient alternative to polyethylene glycol (PEG) for DART positive-ion mass calibration.
- Fomblin Y is a suitable and cost-effective negative-ion reference standard for DART mass spectrometry.
- These identified standards enhance the reliability and scope of DART mass calibration.
More Related Videos
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
09:34Improved Polymerase Chain Reaction-restriction Fragment Length Polymorphism Genotyping of Toxic Pufferfish by Liquid Chromatography/Mass Spectrometry
Published on: September 20, 2016
Related Concept Videos
Tandem Mass Spectrometry
MALDI-TOF Mass Spectrometry
Mass Analyzers: Overview
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Analyzers: Common Types
High-Resolution Mass Spectrometry (HRMS)