Effect of Scatter Limitation Correction with Misregistration between Computed Tomography and Positron Emission
Akihiro Furuta1, Hideo Onishi, Yoshiharu Kangai
1Department of Radiology, Hiroshima City Asa Citizens Hospital.
Purpose:
This study aimed to evaluate the advantage of scatter limitation correction with misregistration between μ-map in the computed tomography attenuation correction and positron emission tomography in PET/CT study.
Methods:
We used torso phantom including simulated tumor and arms phantom. The CT scan was performed by changing the position of arms phantom after PET scan. Arms phantom movement was out-side direction, in-side direction, and top-side direction by 1-12 cm, respectively. The standardized uptake value (SUV) of simulated tumor and background (B.G.) were evaluated for three specific parameters. Two scatter corrections were performed with scatter correction (SC), and scatter limitation correction (SLC).
Results:
The SUVmax of simulated tumor was increased by 2.80% (SC), and 2.78% (SLC) on out-side arms movement. In the SUVmax, SC and SLC were decreased by 28.6%, 9.04% on in-side arms, respectively. SUVmax of the SC, and SLC were increased on top-side arms. The scatter fraction factor (SFF) of SC and SLC were 0.25, 0.25 on out-side 5 cm and were 0.732, 0.391 on in-side 5 cm and were 0.785, 0.434 on top-side 12 cm, respectively.
Conclusion:
SLC improved the overestimation of the SUVmax by SC. However, it is necessary to pay attention, in order not to be improved completely. The finding results indicated that SFF was setting 0.40-0.45 in our institute PET/CT system.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography