Related Experiment Video
Updated: Apr 28, 2026

A Low Cost Setup for Behavioral Audiometry in Rodents
Published on: October 16, 2012
Fixed-Level Frequency Threshold Testing for Ototoxicity Monitoring
Catherine C Rieke1, Odile H Clavier, Lindsay V Allen
11Geisel School of Medicine at Dartmouth, Hanover, New Hampshire, USA; 2Creare, LLC, Hanover, New Hampshire, USA; and 3Walter Reed National Military Medical Center, Bethesda, Maryland, USA.
A new fixed-level frequency threshold (FLFT) test is repeatable and accurate for monitoring high-frequency hearing loss. This rapid, 30-second test aids in detecting ototoxicity and improving patient compliance.
Area of Science:
- Audiology
- Ototoxic monitoring
- High-frequency hearing assessment
Background:
- Ototoxicity often causes high-frequency hearing loss.
- Current monitoring methods can be time-consuming.
- Need for rapid, accurate high-frequency hearing tests.
Purpose of the Study:
- Evaluate the repeatability and accuracy of a Békésy-like, fixed-level frequency threshold (FLFT) technique.
- Determine if FLFT can provide a rapid and effective method for assessing the highest audible frequency.
- Compare FLFT with traditional audiometry in the sensitive region for ototoxicity (SRO).
Main Methods:
- Twenty-nine normal-hearing subjects underwent repeated automated audiometry tests over three weeks.
- Fixed-frequency audiometry identified the highest audible frequency and SRO.
- The FLFT test, administered at 80 dB SPL, used a Békésy-style tracking method.
- Repeatability assessed by root mean square difference; accuracy by difference from fixed-frequency audiometry.
Main Results:
- FLFT showed good repeatability with an average intersession variability of 0.05 octaves.
- No significant learning effect was observed for the FLFT test.
- FLFT accuracy was high, with a mean absolute difference of 0.03 octaves compared to fixed-frequency audiometry.
- FLFT completion time was approximately 30 seconds, significantly faster than traditional methods.
Conclusions:
- The Békésy-style FLFT test is repeatable and accurate for determining the highest audible frequency.
- FLFT demonstrates strong agreement with fixed-frequency audiometry.
- The FLFT test's speed and ease of use make it suitable for quick, accurate, noise-tolerant ototoxicity and high-frequency hearing monitoring.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Therapeutic Drug Monitoring: Drug Analysis Methods