Local activation time sampling density for atrial tachycardia contact mapping: how much is enough?
Steven E Williams1, James L Harrison1, Henry Chubb1
1Division of Imaging Sciences and Biomedical Engineering, King's College London, 4th Floor North Wing, St Thomas' Hospital, 249 Westminster Bridge Road, SE1 7EH London.
Summary
Determining optimal electrode sampling density is crucial for accurate local activation time (LAT) mapping in atrial tachycardia. Higher sampling densities are needed for complex activations and anatomies, with approximately 1.0-1.5 points/cm² recommended for LAT map interpolation.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Imaging
Background:
- Local activation time (LAT) mapping is essential for diagnosing atrial tachycardia.
- While electroanatomic mapping (EAM) systems offer validated accuracy, the impact of electrode sampling density on LAT map reconstruction remains unclear.
Purpose of the Study:
- To investigate the effect of chamber geometry and activation complexity on optimal LAT sampling density.
- To determine the ideal electrode sampling density for accurate LAT map reconstruction in atrial tachycardia.
Main Methods:
- Combined in silico and in vivo approach using atrial monolayers and porcine/human atria.
- Simulated activation patterns at varying densities (0.25-10 points/cm²) and analyzed in simple/complex geometries and activation patterns.
- Defined optimal sampling density as the highest density yielding minimal error reduction in re-interpolated LAT maps.
Main Results:
- Optimal sampling densities varied by tachycardia type in the human left atrium (LA): focal (0.67±0.17), macro-re-entry (1.05±0.32), and localized re-entry (1.23±0.26 points/cm²).
- Increased activation complexity and chamber geometry (porcine right atrium vs. human LA) significantly increased optimal sampling density (P<0.0001 and P=0.0015, respectively).
- In silico simulations showed higher optimal densities for complex activations (re-entry, spiral-wave) compared to focal activation (P<0.0001).
Conclusions:
- Optimal sampling densities can be identified to maximize the diagnostic yield of LAT maps.
- Higher electrode sampling densities are necessary for accurately depicting complex cardiac activation patterns and anatomies.
- An overall optimal sampling density of approximately 1.0-1.5 points/cm² is recommended for LAT map interpolation.