Temporal Artery Temperature Measurement in the Neonate
Mashette E Syrkin-Nikolau1, Karen J Johnson1, Tarah T Colaizy1
1Department of Pediatrics, University of Iowa, Iowa City, Iowa.
American Journal of Perinatology
|April 11, 2017
Summary
An infrared temporal artery thermometer is not more accurate than the axillary thermometer for neonatal temperature measurement. Temporal artery thermometer accuracy in neonates improved with postmenstrual age and varied by bed type.
Area of Science:
- Neonatal care
- Medical device evaluation
- Temperature measurement
Background:
- Accurate temperature monitoring is critical in neonatal care.
- Traditional methods like axillary and rectal thermometry have limitations.
- Novel non-invasive methods require rigorous validation.
Purpose of the Study:
- To compare the accuracy of an infrared temporal artery thermometer against the clinical standard axillary thermometer and the gold standard rectal thermometer in neonatal patients.
- To assess the influence of bed type and postmenstrual age on the accuracy of temporal artery temperature measurements.
Main Methods:
- Prospective study involving 49 neonatal patients.
- Simultaneous measurement of temporal artery (Tta), axillary (Tax), and rectal (Tr) temperatures.
- Statistical analysis comparing Tta and Tax against Tr, considering bed type and postmenstrual age.
Main Results:
- All measurement methods showed significant differences.
- Axillary temperature (Tax) was closer to the gold standard rectal temperature (Tr) than temporal artery temperature (Tta) (mean difference Tr-Tax = 0.21°C vs Tr-Tta = -0.34°C).
- Temporal artery thermometer accuracy improved with increasing postmenstrual age and was less accurate for infants in incubators compared to those in cribs.
Conclusions:
- The infrared temporal artery thermometer is not more accurate than the axillary thermometer for neonatal temperature assessment when compared to the rectal gold standard.
- Temporal artery thermometer accuracy is influenced by environmental factors (bed type) and patient factors (postmenstrual age).
- Further research may be needed to optimize the use of temporal artery thermometers in specific neonatal populations.