Related Experiment Videos
Successful treatment of a cardiac resynchronization therapy nonresponder by identifying lead malpositioning
Karthik Venkatesh Prasad1, Krittapoom Akrawinthawong1, Scott Wayne Ferreira1
1Electrophysiology Services and the Center for Comprehensive Cardiovascular Care (C4), St. Louis University Hospital and School of Medicine, St. Louis, Missouri.
Abstract:
This case describes some of the commonly overlooked device-related issues in patients who have reportedly failed to respond to cardiac resynchronization therapy (CRT). The case demonstrates voltage-dependent right ventricular capture instead of right atrial capture by a subtly malpositioned right atrial lead. CRT therapy failed to improve symptoms of heart failure and the diagnosis of "CRT nonresponder" was made. With a detailed fact-finding approach, the mechanism behind this nonresponse was identified, and the outcome of CRT was significantly improved with rectification of the problems.