Related Experiment Video
Updated: Mar 3, 2026

08:34
Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
21.2K
The Mobius AIRO mobile CT for image-guided proton therapy: Characterization & commissioning
Jasmine A Oliver1, Omar A Zeidan1, Sanford L Meeks1
1Department of Radiation Oncology, UF Health Cancer Center - Orlando Health, Orlando, FL, USA.
Journal of Applied Clinical Medical Physics
|April 25, 2017
Summary
The Mobius AIRO Mobile CT System offers accurate localization for image-guided proton therapy, achieving sub-millimeter precision. Despite slightly lower image quality than other CT systems, its reliability makes it suitable for proton therapy applications.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Imaging Systems
Background:
- The integration of advanced imaging systems is crucial for precise image-guided radiation therapy.
- Mobile CT systems offer potential advantages in workflow and accessibility for various clinical applications.
Purpose of the Study:
- To characterize the Mobius AIRO Mobile CT System for its utility in localization and image-guided proton therapy.
- To evaluate the AIRO system's performance in the context of proton therapy, marking its first known application in this modality.
Main Methods:
- CT images of a Catphan phantom were acquired using the AIRO Mobile CT System and compared with other CT simulators (Philips, Siemens) and a Varian EDGE cone-beam CT.
- Image quality parameters including spatial resolution, contrast, uniformity, and dose were analyzed using DoseLAB software.
- Localization accuracy was assessed using phantoms (MIMI, Stereophan) and compared between 3D and 2D imaging systems.
Main Results:
- The AIRO system demonstrated a spatial resolution of 0.21 lp mm⁻¹, lower than other CT simulators and CBCT systems.
- Significant scaling discrepancies were observed between the AIRO and other systems (AIRO/Siemens: 191.2%, AIRO/Philips: 145.8%).
- Localization accuracy was found to be within 0.6° and 0.5 mm, with maximum shifts of 0.2 mm in the z-direction.
Conclusions:
- The Mobius AIRO Mobile CT System exhibits localization accuracy suitable for image-guided proton therapy, achieving sub-millimeter precision.
- Despite having slightly lower overall image quality compared to conventional CT systems, the AIRO system is deemed accurate and reliable for proton therapy applications.