Agreement between finger plethysmography- and brachial oscillometry-derived blood pressure measurements
Philip D Allan1,2, Terrence O'Donnell1,2, Yu-Chieh Tzeng1,2
1Wellington Medical Technology Group, Department of Surgery & Anaesthesia, University of Otago, Wellington, Wellington, New Zealand.
Purpose:
Blood pressure (BP) is commonly assessed by brachial oscillometry in clinical practice, whereas in physiological studies, finger plethysmography is often employed. This study assessed the limits of agreement between BP metrics obtained from each device.
Methods:
In 96 participants, we simultaneously recorded BP by brachial oscillometry (BP+; Uscom, Sydney, NSW, Australia) and finger plethysmography (Finometer MIDI, MLE1054-V; Finapres Medical Systems B.V., Amsterdam, the Netherlands). Agreement between the two devices was assessed by correlation and Bland-Altman analysis. We assessed average BP differences between the two devices using the criteria of the Association for the Advancement of Medical Instruments (AAMI), which require systolic and diastolic BP differences to be within ≤5 ± 8 (mean ± SD).
Results:
Bland-Altman analysis showed wide limits of agreement (±~17 mmHg or greater) between finger-derived brachial and oscillometric BP. Both systolic and mean BP exhibited positive proportional biases (both P<0·05). Systolic BP differed significantly between devices (7·4 ± 17·7 mmHg, P<0·001), which did not meet the AAMI criteria. No mean bias was observed for diastolic BP (-1·5 ± 8·6 mmHg, P = 0·097), and the SD of ±8·6 mmHg is potentially acceptable given the finger signal may be expected to capture biological variability in BP. Mean BP showed poor concordance (3·7 ± 10·5 mmHg, P<0·001).
Conclusions:
These findings indicate that systolic and mean BP measurements made by brachial oscillometry do not agree with those from finger plethysmography. In contrast, diastolic BP values show acceptable agreement.
More Related Videos
08:42Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure