Carotid atherosclerotic plaques standardised uptake values: software challenges and reproducibility
Nicola Giannotti1, Martin J O'Connell2,3, Shane J Foley4
1Radiography and Diagnostic Imaging, School of Medicine, University College Dublin, Dublin, Ireland. Nicola.giannotti@ucd.ie.
Background:
Positron emission tomography-computed tomography (PET-CT) carotid standardised uptake values (SUV) of 18F-fluorodeoxyglucose (18FDG) have been proposed as an inflammatory biomarker for determining cerebrovascular diseases such as stroke. Consideration of varying methodological approaches and software packages is critical to the calculation of accurate SUVs in cross-sectional and longitudinal patient studies. The aim of this study was to investigate whether or not carotid atherosclerotic plaque SUVs are consistent and reproducible between software packages. 18FDG-PET SUVs of carotids were taken in 101 patients using two different software packages. Quality assurance checks were performed to standardise techniques before commencing the analysis where data from five to seven anatomical sites were measured. A total of ten regions of interest were drawn on each site analysed. Statistical analyses were then performed to compare SUV measurements from the two software packages and to explore reproducibility of measurements. Lastly, the time taken to complete each analysis was measured and compared.
Results:
Statistically significant differences in SUV measurements, between the two software packages, ranging from 9 to 21.8% were found depending on ROI location. In 79% (n = 23) of the ROI locations, the differences between the SUV measurements from each software package were found to be statistically significant. The time taken to perform the analyses and export data from the software packages also varied considerably.
Conclusions:
This study highlights the importance of standardising all aspects of methodological approaches to ensure accuracy and reproducibility. Physicians must be aware that when a PET-CT data set is analysed, subsequent follow-ups must be verified, if possible, with the same software package or cross-calibration between packages should be performed.
More Related Videos
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012
13:45A Method to Study the Correlation Between Local Collagen Structure and Mechanical Properties of Atherosclerotic Plaque Fibrous Tissue
Published on: November 11, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging