Performing Intracochlear Electrocochleography During Cochlear Implantation
Published on: March 8, 2022
5.1K
Cochlear implant electrode sealing techniques and related intracochlear pressure changes.
Ingo Todt1, Julica Utca2, Dania Karimi2
1Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Warenerstr.7, 12683, Berlin, Germany. todt@gmx.net.
Summary
Sealing cochlear implant insertions impacts intracochlear pressure. Pushing tissue in, like with the donut-like fascia ring, significantly increases pressure, which is not recommended for hearing preservation.
Area of Science:
- Otolaryngology
- Biomedical Engineering
- Neurosurgery
Background:
- Cochlear implant electrode insertion requires sealing the cochleostomy or round window to prevent infection and leakage.
- Maintaining low intracochlear pressure is crucial for hearing preservation during cochlear implantation surgery.
Purpose of the Study:
- To investigate the impact of different sealing techniques on intracochlear pressure changes.
- To compare pressure variations associated with various methods of sealing cochlear implant sites.
Main Methods:
- Cochlear implant electrode insertions were simulated in an artificial cochlear model.
- Intracochlear pressure was measured using a micro-pressure sensor in the apical region.
- Four sealing techniques were evaluated: overlay, overlay with pushed-in fascia, donut-like fascia ring, and donut-like fascia ring with pushed-in fascia.
Main Results:
- Statistically significant differences in intracochlear pressure peak amplitudes were observed across sealing techniques.
- The overlay technique resulted in the lowest pressure amplitude (0.14 ± 0.06 mmHg).
- The donut-like fascia ring pushed in technique yielded the highest pressure amplitude (1.79 ± 0.69 mmHg).
Conclusions:
- Sealing techniques for cochlear implant electrode insertion can significantly alter intracochlear pressure.
- Techniques involving pushing sealing tissue into the cochlea are not advisable due to elevated pressure changes.