Validation of plastic scintillation detectors for applications in low-dose-rate brachytherapy
François Therriault-Proulx1, Luc Beaulieu2, Sam Beddar3
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX.
Purpose:
To develop a plastic scintillation detector (PSD) capable of accurately measuring dose around a low-dose-rate (LDR) iodine-125 (125I) radioactive seed as a first step toward in vivo dosimetry for prostate LDR brachytherapy.
Methods:
Using a GEANT4 based Monte Carlo code, photon energy distribution at any position around a realistic 125I source model was obtained. This energy distribution was convolved with the expected energy response from a plastic scintillator and dosimetry accuracy was evaluated. A PSD was constructed and validated in a water phantom for the entire range of clinically relevant positions around an 125I radioactive seed.
Results:
The effect of energy dependence on dosimetry accuracy was shown to be limited, with a maximum relative difference of 1.2% from the calibration condition. A sophisticated approach to account for the energy dependence of PSDs is, therefore, not required if the detector is calibrated using the same model of radioactive seed or a geometrically similar one. The measurements were in good agreement with theoretical models for the entire clinical range.
Conclusions:
This study shows that PSDs can be used for accurate dosimetry in real time around a single 125I seed used in LDR prostate brachytherapy and is promising for clinical applications.