Carbon Dioxide Exposure Resulting From Hood Protective Equipment Used in Joint Arthroplasty Surgery
Suhani Patel1, Janelle M Fine1, Michael J Lim2
1Department of Medicine, University of California, San Diego, California.
Background:
To protect both the surgeon and patient during procedures, hooded protection shields are used during joint arthroplasty procedures. Headache, malaise, and dizziness, consistent with increased carbon dioxide (CO2) exposure, have been anecdotally reported by surgeons using hoods. We hypothesized that increased CO2 concentrations were causing reported symptoms.
Methods:
Six healthy subjects (4 men) donned hooded protection, fan at the highest setting. Arm cycle ergometry at workloads of 12 and 25 watts (W) simulated workloads encountered during arthroplasty. Inspired O2 and CO2 concentrations at the nares were continuously measured at rest, 12 W, and 25 W. At each activity level, the fan was deactivated and the times for CO2 to reach 0.5% and 1.0% were measured.
Results:
At rest, inspired CO2 was 0.14% ± 0.04%. Exercise had significant effect on CO2 compared with rest (0.26% ± 0.08% at 12 W, P = .04; 0.31% ± 0.05% at 25 W, P = .003). Inspired CO2 concentration increased rapidly with fan deactivation, with the time for CO2 to increase to 0.5% and 1.0% after fan deactivation being rapid but variable (0.5%, 12 ± 9 seconds; 1%, 26 ± 15 seconds). Time for CO2 to return below 0.5% after fan reactivation was 20 ± 37 seconds.
Conclusion:
During simulated joint arthroplasty, CO2 remained within Occupational Safety and Health Administration (OSHA) standards with the fan at the highest setting. With fan deactivation, CO2 concentration rapidly exceeds OSHA standards.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...