Background:
The fraction of exhaled nitric oxide (FENO) is a useful marker of asthma control. The FENO measurement with two hand-helded analyzers (NObreath® and NIOX Vero®) may be more affordable, no studies have examined the differences in FENO values measured with those methods in adult.
Methods:
The study population comprised 44 subjects at our outpatient clinic. FeNO values (FENOb and FENOv) were measured by two methods (NObreath® and NIOX Vero®).
Results:
FENOb values were significantly correlated with FENOv (r = 0.911, p < 0.001). However, FENOv values were high compared with FENOb (FENOv = 1.40 × FENOb).
Conclusion:
Differences exist in the values of FENO measured by two hand-helded analyzers: conversion equations are needed to compare the FENO values between these methods.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.