Sci-Thur PM: YIS - 03: Comparing 4D-VMAT, Gated-VMAT and 3D-VMAT in SBRT treatment of lung cancer
Purpose:
To evaluate the treatment plan qualities of 4D-VMAT, gated-VMAT and 3D-VMAT in the treatment of non-small cell lung cancer (NSCLC) in stereotactic body radiation therapy (SBRT).
Methods:
4D-VMAT is a motion compensation strategy that aims to exploit relative target and OAR motion to increase OAR sparing over 3D-VMAT without the long treatment times associated with gated-VMAT. The 4D-VMAT algorithm incorporates the entire patient respiratory cycle and 4D-CT in the optimization process. Resulting treatment plans synchronize the delivery of each MLC aperture to a specific phase of the target motion. Using software developed in Matlab™, SBRT treatment plans for 4D-VMAT, gated-VMAT and 3D-VMAT were generated on 3 patients with NSCLC. Tumour motion ranged from 1.4-3.4 cm. The fractionation scheme was 48Gy in 4 fractions with the GTV receiving 100% of the prescribed dose. For gated-VMAT, the treatment window constrained residual tumour motion to 3 mm or less corresponding to duty cycles of 40-60%. In 3D-VMAT, the ITV was generated by merging the GTV from all phases. A b-spline transformation model was used to register the 4D-CT images and DVHs were calculated from total dose accumulated on the max expiration phase.
Results And Conclusion:
For the majority of OARs, gated-VMAT provided the greatest radiation sparing but significantly extended treatment times (25-35 gantry interruptions/arc). For 3D-VMAT, only 2 patients had clinically acceptable plans that met all the strict dose limits. OAR sparing in 4D-VMAT was comparable to gated-VMAT but with significantly improved delivery efficiency.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...