SU-E-T-413: Dose Verification for Linac-Based SRS Commissioning and Patient Specific QA
A new method using a micro-chamber and solid water phantom simplifies stereotactic radiosurgery (SRS) dose verification. This approach enhances accuracy for both commissioning and patient-specific quality assurance in SRS treatments.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Stereotactic radiosurgery (SRS) program commissioning and quality assurance (QA) traditionally rely on specialized phantoms like the Radiological Physics Center (RPC) Head Neck phantom with TLDs.
- Limited accessibility of the RPC Head Neck phantom restricts routine QA and patient-specific dose verification in many clinics.
Purpose of the Study:
- To propose and validate a simplified SRS dose verification method using widely available clinical instruments.
- To enable both SRS commissioning and patient-specific dose verification.
Main Methods:
- SRS treatment plans were applied to a 14cm solid water phantom with an embedded PTW micro-chamber.
- Treatment couch angles were adjusted to minimize micro-chamber angle dependence and avoid collisions.
- Arc verification plans were recalculated and compared to micro-chamber measurements.
- End-to-end tests using a CyberKnife Head Neck phantom and GAFChromic film served as secondary verification.
Main Results:
- Micro-chamber measurements agreed with calculations within 3% for cone sizes from 10mm to 30mm.
- A 10% underdose was detected with a 5mm cone size, highlighting potential inaccuracies with small cones.
- CyberKnife phantom measurements confirmed dose delivery accuracy within 5%.
Conclusions:
- A straightforward SRS dose verification method employing a micro-chamber and conventional solid water phantom has been successfully presented and validated.
- This accessible approach is valuable for SRS commissioning and ongoing quality assurance.