SU-E-J-56: Localization Error in Gamma Radiosurgery with 3T MR Due to Fiducial Box Chemical Shift from Incorrect
X Ding1, J Bourland1, W Dolesh1
1Wake Forest Univ, Winston-Salem, NC.
Medical Physics
|May 19, 2017
Summary
Using the wrong liquid in gamma radiosurgery (GSR) MRI fiducial boxes causes significant localization errors. MAGIC polymer gel is a versatile tool for GSR quality assurance, detecting shifts up to 3.8mm.
Area of Science:
- Medical Physics
- Radiotherapy Physics
- Medical Imaging
Background:
- Gamma radiosurgery (GSR) relies on accurate target localization using MRI fiducial boxes.
- Incorrect filling materials in fiducial boxes can lead to chemical shift artifacts in MRI.
- These artifacts may cause significant geometric errors in treatment planning.
Purpose of the Study:
- To evaluate localization and positioning errors in GSR due to incorrect liquid filling material (baby oil) in MRI fiducial boxes.
- To assess the utility of a 3D MAGIC polymer gel dosimeter for quality assurance in GSR.
Main Methods:
- Two phantoms (grid and skull) were used with 'baby oil' fiducial markers and a MAGIC gel capsule simulating an intracranial target.
- Phantoms underwent stereotactic MR and CT scanning for GSR treatment planning.
- 3D dose distribution was read out using the MAGIC gel in a 3T MRI scanner; geometric errors were compared to direct measurements.
Main Results:
- A 3.8mm shift in the phase encoding direction was observed for the 'baby oil' fiducial marker in 3T MRI-based GSR planning.
- A 3.7mm shift in Teflon rod position was confirmed by comparing CT and MR images.
- These findings indicate significant geometric inaccuracies caused by the incorrect filling material.
Conclusions:
- Inappropriate filling materials in GSR MRI fiducial boxes can cause substantial target localization and positioning errors.
- The study demonstrates that MAGIC polymer gel is a valuable tool for 3D quality assurance in GSR.
- Accurate material selection for fiducial boxes is critical for safe and effective radiosurgery.