SU-E-T-196: Commissioning for Volumetric Modulated Radiation Therapy on Varian Clinac 21EX
R Yada1,2,3, N Hayashi1,2,3, T Matsunaga1,2,3
1Seirei Hamamatsu General Hospital, Hamamatsu, Shizuoka.
Purpose:
Recently the volumetric arc therapy (VMAT) technology such as RapidArc is widely distributed in Japan. These technologies are normally provided by the high spec linear accelerator such as Trilogy, Novalis Tx, Synergy, et al. The specific DICOM-file is generally used for commissioning of these technologies. On the other hand, we had to apply RapidArc into historic linear accelerator. This title expresses an experience how we performed the commissioning of RapidArc with the old linear accelerator.
Methods:
Two Varian's linear accelerators "Clinac 21EX" equipped with Millenium multi-leaf collimator and a Varian's treatment planning system "Eclipse ver.8.9" were used for this study. The commissioning for RapidArc was performed in energy 4,6,10,15 MV (Max-DR: 250, 600, 400, 600 MU/min). Commissioning procedure composed two categories: the general machine QA for DMLC-IMRT procedure and the specific RapidArc QA procedure. In RapidArc QA procedure, we modified DICOM-file to apply into the potential spec of Clinac 21EX optimally. The specific MLC-motion sequence and the gantry rotation speed were created by the dedicated programs (Shaper and DicomEdit, Varian) for RapidArc QA procedure. Each tolerance value was defied by the data from daily/monthly QA and the paper by Ling et al.
Results:
As the results of the general machine QA procedure, the variance of radiation output during static/dynamic gantry rotation was less than 1%. The deference of fence tests during static/dynamic gantry rotation and RapidArc were less than 1 mm in each. However, the results of the RapidArc QA were worse than the latest machine (especially variable gantry speed) and it was careful to define tolerance level.
Conclusion:
The procedure of commissioning for RapidArc on historic linear accelerator was proposed. Several minor revisions for DICOM-file should be required for suitable commissioning and it may ensure the tolerance limit for gantry/MLC-leaf motion speeds.
More Related Videos
Related Concept Videos
Radiation: Applications
The average...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...